ImPulse
Elite Patient Manual
Figure 11
1. Check the energy level of the battery each time you use the unit.
Refer to “Testing and Replacing Battery” for step-by-step instructions.
2. Check the oxygen cylinder setup and the cylinder contents gauge.
If you need to change cylinders, which includes removal and
reconnection of the ImPulse Elite unit, refer to Oxygen Cylinder
Connection for step-by-step instructions.
3. Make sure the PULSE FLOW / CONTINUOUS FLOW switch is moved
completely right (for pulse flow setting).
4. Set the ON/OFF switch to the ON position.
All three lights illuminate for less than 1 second. The
ImPulse Elite unit automatically calibrates and reaches
sensitivity within approximately 15 seconds. Then the
green light shuts off, and the unit emits a short burst of
oxygen. This indicates ImPulse Elite is ready for use.
5. Connect your cannula to the oxygen outlet. (You may
use up to a 25 ft [7.5 m] cannula, if desired.) (Figure 11.)
6. Turn the cylinder valve slowly one full turn counter-
clockwise to begin oxygen flow.
7. Make sure the cylinder contents gauge on the ImPulse Elite unit registers
within the green shaded/green zoned areas.
This indicates an adequate oxygen supply in the cylinder. (If the cylinder
contents gauge indicates the red REFILL zone, the
cylinder needs to be replaced.)
8. Position the nasal cannula with the prongs inserted
into your nostrils.
9. Set the flow selector between 1 and 6 depending on
your prescribed dosage (Figure 12).
Figure 12
11